Troubleshooting

Symbol

Possible Cause

What To Do

 

 

 

 

 

 

 

The meter detected

1.Repeat test with new

 

a temperature

test strip. Make sure

 

problem.

meter and test strip are

 

 

at room temperature.

 

 

• Do not apply heat or

 

 

cold to meter. Operating

 

 

temperature is 18°–30°C.

 

 

2.If you continue to get

 

 

any error message,

 

 

obtain a laboratory

 

 

result, and

 

 

• Compare color of test

 

 

strip confirmation dot to

 

 

Color Chart on test strip

 

 

bottle to see if glucose

 

 

level is very high or very

 

 

low.

 

 

• Take appropriate steps

 

 

to monitor patient and

 

 

provide medical care

 

 

while awaiting

 

 

laboratory result.

 

 

3.Contact LifeScan for help.

 

 

 

LifeScan Healthcare Professional Line: 1 800 524-7226