Symbol

Possible Cause

What To Do

 

 

 

 

 

 

 

The meter detected

1.Clean meter, especially

 

a problem with the

gray area inside test

 

optical system.

strip holder, but

NOTE: Before cleaning, compare color of test strip confirmation dot to Color Chart on test strip bottle. If glucose level is very high or very low, take appropriate steps to monitor patient and provide medical care.

• Repeat test with new test strip.

• Make sure test strip holder is fully inserted into meter.

• Wait for “Insert test strip” symbol to appear on display before inserting test strip.

2.If you continue to get any error message, obtain a laboratory result. Take appropriate steps to monitor patient and provide medical care while awaiting laboratory result.

3.Contact LifeScan for help.

LifeScan Healthcare Professional Line: 1 800 524-7226