Specifications

Thermistor Range (10 µA)

0.0 to 450.0 KΩ

 

 

Thermistor Range (100 µA)

0.0 to 45.0 KΩ

 

 

Usable Thermistor Range

25 to 450,000 Ω, typical

 

 

Thermistor Resistance Resolution

0.1 kΩ

(10 µA)

 

Thermistor Resistance Resolution

0.01 kΩ

(100 µA)

 

 

 

Thermistor Resistance Accuracy

+0.05% of FS

 

 

AD590 Reverse Bias

8 Volts

 

 

LM335 Bias

0.6 mA

 

 

User Calibration

 

Thermistor

IC Sensor

Steinhart-Hart equation (2 constants)

2-point

Measurement (Display

Display Type

4-digit LED

 

 

TE Current Range

-4.00 to 4.00 Amps

 

 

TE Current Resolution

0.01 Amps

 

 

TE Current Accuracy

+0.03 Amps

 

 

Temperature Range

-99.9 to 999.9oC

Temperature Resolution

0.1oC

Temperature Accuracy6

0oC: +0.6oC, typical

 

20oC: +0.3oC, typical

 

30oC: +0.4oC, typical

 

50oC: +0.6oC, typical

6Accuracy figures quoted are typical for a calibrated 10K thermistor; accuracy figures are relative to calibration standard and are dependent on the user-defined configuration of the instrument; variation from typical value is largely due to uncertainty in thermistor calibration