MAINTENANCE AND TR OUBLESHOOTING

Troubleshooting

Troubleshooting

C H A P T E R 3

This section is a guide to troubleshooting the LDT-5525 Temperature Controller. Some of the more common symptoms are listed here, and the appropriate troubleshooting actions are given. We recommend that the user start at the beginning of this guide. Read the symptom descriptions, and follow the steps for the corrective actions which apply. If you encounter problems which are beyond the scope of this guide, contact your ILX Lightwave representative.

 

 

 

 

 

Symptom

Causes and Corrective Actions

 

 

 

 

 

 

LDT-5525 Series unit will not

Check AC power line voltage and power cord connection

 

 

power up

 

 

 

 

 

 

 

Power on, but display is frozen,

This may occur if the unit loses power (AC line) briefly. Turn the

 

 

switches don’t work

power switch off and on again to restart

 

 

 

 

 

 

Power on, but no TE current

If TE OPEN indicator is lit, check the load connections and then

 

 

output

try again

 

 

 

Check that pins 14 and 15 of the output connector are not

 

 

 

connected (see Chapter 2: Booster Operation)

 

 

Power on, but measured ITE

Check that pins 14 and 15 of the output connector are not

 

 

current is always about 0.0 A

connected (see Chapter 2: Booster Operation)

 

 

 

 

 

 

Power on, but temperature is

If SENSOR OPEN indicator is lit, check the sensor connections

 

 

not controlled

and then try again

 

 

 

Check that the back panel SENSOR SWITCH position is set to the

 

 

 

proper sensor type

 

 

 

Check that C1 and C2 are correct values for your sensor

 

 

 

Check that the GAIN setting is not too low and that the I LIMIT

 

 

 

value is not too low for your thermal load

 