Lightwave Communications LDT-5525 manual AD590 Sensor Calibration

Models: LDT-5525

1 56
Download 56 pages 22.8 Kb
Page 36
Image 36
AD590 Sensor Calibration

C H A P T E R 3

MAINTENANC E AND TR OUBLESHOOTING

Calibration Adjustments

AD590 Sensor Calibration

The following procedure is for calibrating the AD590 sensor measurement so that the temperature measurement will be as accurate as possible. This procedure calibrates the current measurement of the AD590. This procedure does not calibrate C1 and C2. For information on calibrating the AD590 sensor, see Appendix C.

Calibration may be aborted by pressing the OUTPUT switch.

aSet the SENSOR SELECT switch (back panel) to the AD590 position. Set C1 to 0.00, C2 to 1.00.

bConnect a precision 20 Kmetal film resistor and a precision ammeter in series at the sensor input of the LDT-5525 Temperature Controller.

cEnter the sensor calibration mode by pushing the MODE and SENSOR CAL switches at the same time. After this, the TEC display will indicate sensor reference current in A. Wait for three seconds for the measurement to settle.

dPress and hold in the ENBL switch and turn the ADJUST knob until the display indicates the same current as shown on the precision ammeter.

eRelease the ENBL and wait for the VIEW SET indicator (LED) to be unlit. Replace the 20 Kresistor with a 10 Kmetal film resistor. Wait for three seconds, then repeat Step d using the 10 Kresistor.

Ten seconds after the ENBL switch is released, the LDT-5525 Temperature Controller will store the calibration data in non-volatile memory.

fPress the DISPLAY switch three times to rotate the display back to the SENSOR mode. After calibration, I LIMIT will be automatically set to 4.00 Amps. Reset the I LIMIT to the desired value.

22 LDT-5525

Page 36
Image 36
Lightwave Communications LDT-5525 manual AD590 Sensor Calibration