Recommended Equipment

C H A P T E R 3

MAINTENANC E AND TR OUBLESHOOTING

Calibration Overview

Recommended Equipment

Recommended test equipment for calibrating the LDT-5525 Temperature Controller is listed in Table 3.1. Equipment other than that shown in the table may be used if the specifications meet or exceed those listed.

Table 3.1

DESCRIPTION

MFG / MODEL

SPECIFICATION

 

 

 

DMM

HP 3457A

DC Amps (@ 1.0A): +0.02%

 

 

Resistance (@ 10): 0.02%

 

 

0.1 A or 0.1 mW resolution

Resistors

Metal Film

15 K(for ITE calibration)

 

 

4

Kand 40 K(for 100 mA calibration)

 

 

4

Kand 400 K(for 10 mA calibration)

 

 

4

Kand 10 K(for LM335 calibration)

 

 

10 Kand 20 K(for AD590 calibration)

 

High Power

1

, 20W, low TCR (for ITE calibration)

 

 

 

 

Environmental Conditions

Calibrate this instrument under laboratory conditions. We recommend calibration at 23oC ± 1.0oC. When necessary, however, the LDT-5525 Temperature Controller may be calibrated at its intended use temperature if this is within the specified operating temperature range of 0 to 40oC.

Warm Up

The LDT-5525 Temperature Controller should be allowed to warm up for at least 1 hour before calibration.