Net Safety Monitoring Inc

Calibration Failure

If the calibration procedure fails, the display shows Fail Cal , the Status LED alternates Red and Green flashes and the analog output changes back and forth from 3.0 to 3.3 mA.

The unit remains in a failed state until manually reset. After the Manual Reset, the unit will return to normal operation based on previous calibration values. Refer to "Manual Reset" on page 17 for instructions.

STEP 6 — MONITOR

REVIEW RELAY SETTINGS

This is a read-only mode; changes cannot be made.

Step 1: Press and hold the Calibration Button or the magnet to the Reed Switch to enter the Main Menu; wait for the countdown, from 10 to 0, to end.

Step 2: When Review Relay Settings displays press the Calibration Button or use the Reed Switch to select.

The flashing YES confirms the selection.

Step 3: The Fault Alarm is fixed (Energized / Non-Latching) and displays first. Then the Low Alarm level, coil and latch status display, followed by the High Alarm level, coil and latch status.

Step 4: At this point, the option to Set Relay Options YES? is displayed.

MLP-A/AR/AD-ST1400-25

14

Page 19
Image 19
Millennium Enterprises AD-ST1400-25, MLP-A/AR user manual Monitor, Calibration Failure, Review Relay Settings