Configuration

Configuration

3

 

 

Configuration through Software

DaqBoard/500 Series boards have no hardware jumpers or switches. All data-acquisition settings such as analog input, data collection rates, input voltage range, and operating modes are configured through application software. A DaqView software driver provides an application level software interface to Windows NT, 2000, and XP. Software packages such as LabVIEW can also be used. These packages configure and collect, or output, acquisition data in a GUI based interface.

Analog Input Configuration

The analog inputs are impedance buffered and drive a differential gain amplifier that can be referenced in a number of ways allowing for Single-Ended or Differential programmable input configurations.

A “1 to 176” element channel configuration RAM is provided to allow each ADC channel to be set with a different Gain, Range, Thermocouple Type and Input Configuration selection combination.

ADC Ranges

The analog inputs may be configured for either ±10 V bipolar or 0 to10 V unipolar operation. The input range is programmable on a channel by channel basis in a 176-element channel configuration RAM. Note that the range selection also applies to expansion channels.

The programmable gain circuitry must also be taken into account in defining the usable error free input range. The boards provide a wide range of programmable ranges and resolutions. The following tables indicate the maximum resolution under different conditions. Note that resolution is not accuracy.

Resolution defines the minimum definable voltage increment. Absolute DC accuracy and relative accuracy defines exactly how close the reading will be to the actual voltage input.

ADC INPUT

DaqBoard/500 Series Range Resolution

Bipolar

 

Programmable

 

 

Full Scale

 

 

Microvolt

 

 

 

 

 

 

 

 

Gain

 

 

Range

 

 

Resolution

 

 

x1

 

 

 

 

310.140 V/bit

 

 

± 10.00 V

 

 

 

x2

 

 

± 5.00 V

 

 

155.070 V/bit

 

 

x4

 

 

± 2.50 V

 

 

77.535 V/bit

 

 

x8

 

 

± 1.25 V

 

 

38.768 V/bit

 

Unipolar

 

 

 

 

 

 

 

Programmable

 

 

Full Scale

 

 

Microvolt

 

 

Gain

 

 

Range

 

 

Resolution

 

 

x1

 

 

0 to 10.00 V

 

 

155.070 V/bit

 

 

 

 

 

 

x2

 

 

0 to 5.00 V

 

 

77.535 V/bit

 

x4

 

 

0 to 2.50 V

 

 

38.768 V/bit

 

 

x8

 

 

0 to 1.25 V

 

 

19.384 V/bit

 

DaqBoard/500 Series

879095

Configuration 3-1

Page 27