RayTek MI Miniature Infrared Sensor Connecting of Cables to the Electronic Box

Models: MI Miniature Infrared Sensor

1 95
Download 95 pages 3.96 Kb
Page 25
Image 25

Installation

Preinstalled cable to sensor head

Output signal and power connector block

Cable that has to be installed by the user

Figure 6: Connecting of Cables to the Electronic Box

5.Put the following on the cable (as shown in the figure above): the cap (1), the plastic compression fitting (2), the rubber washer (3) and one of the metal washers (4).

6.Spread the cable shield (5) and then slip the second metal washer (4) on the cable. Note that the shield must make good contact to both metal washers.

7.Slip the wires (6) into the electronic box far enough to connect to the power and output terminals.

8.Screw the cap (1) into the electronics box. Tighten snuggly. Do not over tighten.

9.Connect the wires (6) to the power and output terminals on the printed circuit board.

MI

17

Page 25
Image 25
RayTek MI Miniature Infrared Sensor operating instructions Connecting of Cables to the Electronic Box