Programming Guide

There are eight entries possible for emissivity setting (1) and a related set point (threshold) (2). To be able to write or read these values, use the following commands:

EP=2 set pointer for table entry, e.g. to line 2 (3) RV=0.600 set the emissivity value for line 2 to 0.600 (4)

SV=220.0 set the set point (threshold) for line 2 to 220.0 (5)

1 2

3

4

5

Figure 46: Table for Emissivity and Set Points

To activate these emissivity settings, you need to have the 3 external inputs (FTC) connected. According to the digital combination on the FTC wires, one of the table entries will be activated, see section 5.4.2 Emissivity Setting via Digital Selection on page 25.

MI

71

Page 79
Image 79
RayTek MI Miniature Infrared Sensor operating instructions Table for Emissivity and Set Points