Maintenance

9.3 Sensing Head Exchange

Sensing heads and electronic boxes can only be interchanged in accordance to the following table!

 

MID02

MIC02

MID10

MIC10

MIH10

MID20

MIC20

MIH20

MID02

x

x

x

x

 

 

 

 

MIC02

x

x

x

x

 

 

 

 

MID10

x

x

x

x

 

 

 

 

MIC10

x

x

x

x

 

 

 

 

MIH10

 

 

 

 

x

 

 

 

MID20

 

 

 

 

 

x

 

 

MIC20

 

 

 

 

 

 

x

 

MIH20

 

 

 

 

 

 

 

x

The head exchange requires to type in the new sensing head calibration data printed on the cable as follows:

1.To exchange the sensing head, disconnect the power of the unit.

2.Connect the wires for the new sensing head according to the color description on the printed circuit board.

3.Switch the power for the unit to ON.

4.Press simultaneously the <Mode/Down/Up> buttons.

5.Four characters appear in the display (former values). Type in the new designator (A) using the <Down/Up> buttons. Press the <Mode> button.

6.The second block of four characters appears in the display (former values). Type in the new designator (B) using the

MI

63

Page 71
Image 71