Installation
5.3.1 Signal Output
Power + | Power – Signal | Signal |
| Ground | Output |
Figure 8: Wiring of the Signal Output (mA or V)
The signal output can be configured either as current or as voltage output.
The minimum load impedance for the 0 to 5 V output must be 100 kΩ.
The maximum current loop impedance for the 0/4 to 20 mA output can be 500 Ω, and the power supply and loop impedance must be matched as shown below.
Max. Loop Impedance [Ω]
Max. Loop Impedance
Power
Supply
[V]
Figure 9: Max. Loop Impedance depending on Power Supply
MI | 19 |