Programming Guide

11.7 Multiple Units (RS485 Multidrop Mode)

Up to 32 units can be connected within a RS485 network, see section

5.6Installing of Multiple Sensors via RS485 on page 31. To direct a command to one unit among the 32 possible, it is necessary to „address“ a command. Therefore, a 3‐digit number is set prior the command. The 3‐digit number is determined between 001 and 032. Exception: Broadcast command.

If a command is transferred, starting with the 3‐digit number 000, all units (with addresses from 001 to 032) connected will get this command – without to send an answer.

Note: A unit with the address 000 is a single unit and not in multidrop mode.

XA=024 will set address to 24 (unit must not be in multidrop mode)

Changing an address:

(e.g. the address is change from 24 to 17)

command

answer

„017?E“

„017E0.950“

„017XA=024“

“017XA024” setting of a new address

„024?E“

„024E0.950“

Example for the broadcast command:

command

answer

“024?E”

“024E0.950”

“000E=0.5”

will be executed from all units, no answer

“024?E”

“024E0.500”

“012?E”

“012E0.500”