Programming Guide
11.4.3 Post Processing
The following parameters can be set to determine the post processing mode, see section 6.4 Post Processing on page 38.
P=5  | peak hold, hold time: 5 s  | 
F=12.5  | valley hold, hold time: 12.5 s  | 
G=10  | averaging, average time (90%): 10 s  | 
XY=3  | advanced peak hold, hysteresis: 3 K  | 
XY=‐2  | advanced valley hold, hysteresis: 2 K  | 
  | Advanced Peak/Valley Hold with Averaging:  | 
C=250  | threshold: 250°C  | 
AA=15  | averaging time (90%): 15 s  | 
11.5 Dynamic Data
All temperature related information is calculated 128 times a second. To request the dynamic data, following commands are available:
?T  | target temperature  | 
?I  | internal temperature of the sensing head  | 
?XJ  | internal temperature of electronics housing  | 
?Q  | energy value of the infra‐red temperature  | 
?XT | trigger set point (active/inactive) for the FTC3 input  | 
To check for resets (e.g. power shut down) use the command XI. Notice, after a reset the unit is new initialized.
?XI | asks for the reset status  | 
!XI0  | no reset occurred  | 
!XI1  | a reset occurred, new initialization of the unit  | 
XI=0  | sets the reset status back to 0  | 