Programming Guide

11.4.3 Post Processing

The following parameters can be set to determine the post processing mode, see section 6.4 Post Processing on page 38.

P=5

peak hold, hold time: 5 s

F=12.5

valley hold, hold time: 12.5 s

G=10

averaging, average time (90%): 10 s

XY=3

advanced peak hold, hysteresis: 3 K

XY=‐2

advanced valley hold, hysteresis: 2 K

 

Advanced Peak/Valley Hold with Averaging:

C=250

threshold: 250°C

AA=15

averaging time (90%): 15 s

11.5 Dynamic Data

All temperature related information is calculated 128 times a second. To request the dynamic data, following commands are available:

?T

target temperature

?I

internal temperature of the sensing head

?XJ

internal temperature of electronics housing

?Q

energy value of the infra‐red temperature

?XT

trigger set point (active/inactive) for the FTC3 input

To check for resets (e.g. power shut down) use the command XI. Notice, after a reset the unit is new initialized.

?XI

asks for the reset status

!XI0

no reset occurred

!XI1

a reset occurred, new initialization of the unit

XI=0

sets the reset status back to 0

72

MI

Page 80
Image 80
RayTek MI Miniature Infrared Sensor operating instructions Dynamic Data, Post Processing