Programming Guide
11.2 General Command Structure
Requesting a parameter (Poll Mode)
?ECR | “?“ is the command for “Request“  | 
  | “E“ is the parameter requested  | 
  | “CR“ (carriage return, 0Dh) is closing the request.  | 
  | Remark: It is possible to close with “CR“ “LF“, 0Dh,  | 
  | 0Ah, but not necessary.  | 
Setting a parameter (Poll Mode)
The parameter will be stored into the device EEPROM. E=0.975CR “E“ is the parameter to be set
“=“ is the command for “set a parameter“ “0.975“ is the value for the parameter
“CR“ (carriage return, 0Dh) is closing the request Remark: It is possible to close with “CR“ “LF“, 0Dh, 0Ah, but not necessary.
Setting a parameter without writing into the EEPROM (Poll Mode)
This function is for test purposes only. E#0.975CR “E“ is the parameter to be set
“#“ is the command for “set parameter without writing into the EEPROM“
“0.975“ is the value for the parameter
“CR“ (carriage return, 0Dh) is closing the request. Remark: It is possible to close with “CR“ “LF“, 0Dh, 0Ah, but not necessary.
Device response format:
!E0.975CRLF “!“ is the parameter for “Answer“
“E“ is the parameter
“0.975“ is the value for the parameter “CR“ „LF“ (0Dh 0Ah) is closing the answer.
68  | MI  |