Problem | Possible Cause | Solution |
|
|
|
| There is excessive leak. | Adjust position of mask and |
|
| headgear. |
|
| Caps may be missing from ports |
|
| on mask. Replace them. |
|
| Air tubing not connected |
|
| properly. Connect firmly at both |
|
| ends. |
|
| Air tubing kinked or punctured. |
|
| Straighten or replace. |
|
|
|
S8 AutoSet Spirit unit | SmartStart/Stop is disabled. | Enable SmartStart/Stop. |
does not stop when |
|
|
you remove your |
|
|
mask. |
|
|
|
|
|
SmartStart is enabled | Incompatible humidifier or mask | Use only equipment as |
but the flow generator | system being used. | recommended and supplied by |
does not stop |
| ResMed. |
automatically when |
|
|
you remove your |
|
|
mask. |
|
|
Pressure rises | Talking, coughing, voluntarily |
inappropriately. | |
| over in bed), or intentionally |
| breathing in an unusual manner. |
Avoid talking with a nasal mask on, and breathe as normally as possible.
Mask cushion is buzzing against the skin. Cushion seated incorrectly.
Cushion seated incorrectly causing excessive leak.
Air tubing is kinked or punctured.
Humidifier control dial set too high, resulting in accumulation of water in the air tubing.
Adjust the headgear.
Adjust headgear or
Straighten or replace air tubing.
Turn humidifier control dial down and empty the water from the air tubing.
32