Solid State Logic 82S6XL090A Appendix D - Calibration Information, Output Stage, Level Trim, Page

Models: 82S6XL090A

1 30
Download 30 pages 38.61 Kb
Page 25
Image 25
Appendix D - Calibration Information

Appendix

Appendix D - Calibration Information

The XLogic E Signature Channel unit is factory calibrated and should only need calibration if a potentiometer or other component has been replaced or if it is suspected that there is a problem with calibration.

In all of the following instructions it is assumed that the lid has been removed and that power has been applied. It is also assumed that unless otherwise specified, all switches are released and all potentiometers are at unity, minimum or indent position as appropriate. The required accuracy for each adjustment will be specified along with the target value.

All level and distortion measurements should be made with audio-band 20Hz to 20kHz filters unless otherwise specified. All adjustments are made on the main (629612) card.

Line Input

Equipment Required:

Calibrated audio oscillator and audio level meter

Test Signal:

1kHz sinewave @ +6dBu

Input and Output:

Oscillator to Line Input and TP46 on the 629612 card to the audio level meter

Unit Setup:

Set the Line Gain to indent (0dB)

Level Trim

 

Adjustment:

1. Adjust VR2 for 0dBu ±0.05dB.

Output Stage

Equipment Required:

 

Calibrated audio oscillator, audio level meter and dc. volt meter. A ‘balance’

 

 

adaptor (see below) will be required for the output balance adjustment.

Test Signal:

 

1kHz sinewave @ 0dBu

Input and Output:

 

Oscillator to Line Input and Output to the audio level meter

Unit Setup:

1.

Set both Line Gain and Output Gain to indent (0dB).

 

2.

Press LINE switch IN.

Unity Gain Trim

 

 

Adjustment:

1.

Adjust VR28 for 0dBu ±0.05dB.

DC Offset Trim

 

 

Adjustment:

1.

Turn Oscillator off.

 

2.

Measure on TP54 and adjust VR29 for 0V ±1mV.

 

3.

Measure on TP59 and adjust VR30 for 0V ±1mV.

Output Balance

 

 

Input and Output:

 

Signal source as before, connect the Output to the audio level meter via the

 

 

‘balance’ adaptor

Adjustment:

1.

Adjust VR31 for minimum level (< 55dBr).

‘Balance’ Adaptor

For the output balance adjustment, a ‘balance’ adaptor such as that illustrated here will be required. This adaptor consists of a pair of close tolerance resistors in an in-line cable and is used to sum together a balanced output in order to correctly adjust the level balance of the measured output; perfect balance should result in complete signal cancellation.

 

1

 

3

 

 

0V

 

 

 

 

 

1

 

 

 

From unit

2

5K01**

 

 

 

+

 

 

 

To measuring 2

under test

 

1

2

+

 

3

5K01**

 

equipment

 

 

 

1 Level Trim 0V

Note 1. Resistor tolerence should ideally be 0.01%

2.Absolute level measured will depend upon the input impedence of the measuring equipment.

Page 21

Page 25
Image 25
Solid State Logic 82S6XL090A Appendix D - Calibration Information, Output Stage, Level Trim, Unity Gain Trim, Line Input