Functional Check
18 P6330 3.5 GHz Differential Probe Instruction Manual
NOTE. If your instrument supports probe calibration routines, now is
a good time to perform them.
6. Disconnect the probe from the PROBE COMPENSATION
connector and connect the two KlipChips together.
7. Withtheprobeoffsetset to 0.0 V, the oscilloscope display should
be at the ground reference.
8. Setthe oscilloscopevolts/divisionto 500mV.
9. Adjust the probe offset. The displayed waveform should vary
between approximately +1.0 V and --1.0V.