Adjustments
P6330 3.5 GHz Differential Probe Instruction Manual 51
-- +
Powersupply
1103
CH1
input
CH1
output
FemaleBNC-to-male
bananaadapter
50Ωprecision
terminator
50ΩBNC cable
Y-leadadapter
Black( --)
Digital multimeter Squarepin adapter
P6330
Red(+ )
Digital multimeter
KlipChip
adapters
Figure 25: Offsetzero and DC CMRR setup
6. Set the input voltage on the DC source to 3.0 V,±10 mV.
7. Adjust the CMRR control in the probe for an output equal to the
valuemeasuredinstep4onpage50,±0.1mV.
8. Set the input voltage on the DC source to approximately +3.0 V.
RecordtheactualvoltageasVin1.
9. Record the output voltage as Vout1.
10.Set the i nput voltage on the DC source to approximately --3.0 V.
RecordtheactualvoltageasVin2.
11. Record the output voltage as Vout2.