P6330 3.5 GHz Differential Probe Instruction Manual 37
Performance Verification
Use the following procedures to verify specifications of the P6330
probe. Before beginning these procedures, refer to page 46 and
photocopy the test record, and use it to record the performance test
results. The recommended calibration interva l isone ye ar.
Theseprocedurestest the followingspecifications:
HOutput offset voltage
HDCgainaccuracy
HRisetime

Equipment Required

Refer to Table 7 for a list of the equipment required to verify the
performance of your probe.
Table7: Equipm ent required forperformance verification
Item description Performance requirement Recommended example
High Speed Sampling
Oscilloscope
10 GHz bandwidth TektronixTDS8000 with
80E04 moduleor 11800 with
SD24 module
Probe CalibrationFixture See page 39 067-0419-00
Probe Positioner TektronixPPM203B
Power Supply TEKPROBE interface Tektronix1103
Power Supply 5.0 VDC at 1 mA Tektronix PS280
DMM (2), with leads 0.1 mV resolution Fluke87 or equivalent
Feedthrough Termination BNC,50 ±0.05 011-0129-00
Coaxialcables (2) Male-to-Male SMA, 20 in 174-1427-00
Coaxialcable Male-to-Male BNC, 50 012-1342-00