Command Descriptions
2–110 TDS Family Oscilloscope Programmer Manual
:DIAg : ?
LOG
RESUlt
<QString> in the following format:
<Status>,<Module name>[,<Status>,<Module name>...]
DIAG:RESULT:LOG?
might return :DIAG:RESULT:LOG "pass--Processor,pass--Display,
pass--FP/Proc Interface,FAIL--Front Panel"
DIAg:SELect:ACQUISition (No Query Form)
Selects the acquisition system test sequence that will be run when the
DIAg:STATE EXECUte command is sent. This command is equivalent to setting
Area in the Utility menu when System is set to Diag/Err.
Calibration and Diagnostic
DIAg:SELect:ACQUISition ALL
:SELect
DIAg : ACQUISition <Space> ALL
ALL selects functional, memory, and register tests.
DIAg:SELect:ALL (No Query Form)
Specifies that all system test sequences will be run when the DIAg:STATE
EXECUte command is sent. This command is equivalent to setting Area in the
Utility menu when System is set to Diag/Err.
Calibration and Diagnostic
DIAg:SELect:ALL ALL
:SELect
DIAg : <Space> ALLALL
ALL selects functional, memory, and register tests for the acquisition, processor
and display systems, and self diagnostics for the front panel.