Command Descriptions
2–354 TDS Family Oscilloscope Programmer Manual
WFMPre:<wfm>:NR_Pt
Sets or queries the number of points that are in the transmitted waveform record.
This value is ignored on input.
In extended-acquisition-length mode, it will set or query the number of points
that are in the transmitted the acquisition record.
DATa:DESTination
Waveform
 

 
 

 is the number of data points. If DATa:WIDth is 2 then there are twice as
many bytes.
 = 0 means that the waveform record is of an unspecified length.

might return  as the number of data points in the waveform record trans-
ferred from channel 1.
WFMPre:<wfm>:PT_Fmt
Selects the data point format for the first ordered waveform as selected by the
DATa:SOUrce command. On input <wfm> always defaults to the reference
location specified by DATa:DESTination regardless of what is sent.
Waveform
DATa:DESTination
 !"
Related Commands
Group
Syntax
Arguments
Examples
Group
Related Commands
Syntax