Command Descriptions
TDS Family Oscilloscope Programmer Manual 2–233
LOW
?
<Space> <NR3>
MEASUrement : REFLevel :ABSolute :
<NR3> is the low reference level, in volts. The default is 0.0 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:LOW?
might return 0.0E+0 as the low reference level.
MEASUrement:REFLevel:ABSolute:MID
Sets or queries the mid reference level, and is the 50% reference level when
MEASUrement:REFLevel:METHod is set to ABSolute. This command is
equivalent to setting the Reference Levels in the Measure menu.
Measurement
MEASUrement:REFLevel:ABSolute:MID <NR3>
MEASUrement:REFLevel:ABSolute:MID?
MID
?
<Space> <NR3>
MEASUrement :REFLevel :ABSolute :
<NR3> is the mid reference level, in volts. The default is 0.0 V.
MEASUREMENT:REFLEVEL:ABSOLUTE:MID .71
sets the mid reference level to .71 volts.
Arguments
Examples
Group
Syntax
Arguments
Examples