Command Descriptions
2–248 TDS Family Oscilloscope Programmer Manual
*RCL 3
restores the digitizing oscilloscope from a copy of the settings stored in memory
location 3.
RECAll:ACQDATA (No Query Form)
TDS 500C & 700C Only
Replaces the indicated channel’s live acquisition data with that saved in the
indicated file.
Save and Recall
RECAll:ACQDATA <file-path>, CH<x>
<file path>
<Space>
RECAll :ACQDATA CH <x>
,
<file path> is the location in mass storage memory where the setup will be
recalled from.
<file path> is a quoted string that defines the file name and path. Input the file
path using the form <drive>/<dir>/<filename>. <drive> and one or more
<dir>s are optional. If you do not specify them, the TDS will read the file from
the current directory. <filename> stands for a filename of up to 8 characters
followed by a period (“.”) and any 3-character extension. Do not use wild card
characters.
The current directory refers to the name of a directory as returned by the
FILESystem:CWD command.
CH <x> is an input channel.
RECALL:ACQDATA hd0:/MYFILE.WF1",CH2
replaces CH2’s data with that stored in hd0:/MYFILE.WF1.
RECAll:SETUp (No Query Form)
Restores a stored or factory front-panel setup of the digitizing oscilloscope. This
command is equivalent to selecting Recall Saved Setup or Recall Factory
Setup or Recall Current Setup in the Save/Recall Setup menu.
Save and Recall
Examples
Group
Syntax
Arguments
Examples
Group