Command Descriptions
2–244 TDS Family Oscilloscope Programmer Manual
complete discussion of the use of these registers and the output queue, see page
3–1.
Status and Error
BUSY?, *WAI



The *OPC command allows you to synchronize the operation of the digitizing
oscilloscope with your application program. Synchronization methods are
described starting on page 3–7. Table 2–33 shows commands that generate an
Operation Complete message.
Table 2–33: Commands that Generate an Operation Complete Message
Operation Command
Automatic scope adjustment AUTOSet EXECute
Internal self-calibration *CAL
Single sequence acquisition ACQuire:STATE ON or
ACQuire:STATE RUN
(when ACQuire:STOPAfter is set to
SEQuence)
Hardcopy output HARDCopy STARt
*OPT? (Query Only)
Not on TDS 400A
Returns a list of options installed in your digitizing oscilloscope.
Status and Error

Group
Related Commands
Syntax
Group
Syntax