Command Descriptions
2–54 TDS Family Oscilloscope Programmer Manual

might return , indicating that repetitive signal acquisition mode is on.
ACQuire:STATE
Starts or stops acquisitions. This is the equivalent of pressing the front-panel
RUN/STOP button. If ACQuire:STOPAfter is set to SEQuence, other signal
events may also stop acquisition.
Acquisition
ACQuire:NUMACq?, ACQuire:REPEt, ACQuire:STOPAfter
           

 






 or  or  = 0 stops acquisitions.
 or  or  0 starts acquisition and display of waveforms. If the
command was issued in the middle of an acquisition sequence (for instance
averaging or enveloping),  restarts the sequence, discarding any data
accumulated prior to the . It also resets the number of acquisitions.
 
starts acquisition of waveform data and resets the number of acquisitions count
(NUMACQ) to zero.

returns either or , depending on whether the acquisition system is running.
Group
Related Commands
Syntax
Arguments
Examples