Quickstart Guide

Trace Oxygen Analyzer

 

 

BASIC OPERATION

Once the analyzer has been installed, it can be configured for your application. Prior to using the instrument you must:

Calibrate the instrument

Define the analysis ranges and choose whether to use autoranging or a fixed range.

Set alarm setpoints and modes of alarm operation.

Additional configuration parameters are available to tailor the instrument to your particular application. Refer to the Instruction Manual for specific information on setting of additional system parameters.

Default Configuration

The default analyzer configuration is as follows:

Ranges: LO = 1 ppm, MED = 10 ppm, HI = 100 ppm.

Auto Ranging: ON

Alarm Relays: Defeated, 10.00 ppm, HI, Not failsafe, Not latching. Zero: Auto, every 0 days at 0 hours.

Span: Auto, at 000,008.00 ppm, every 0 days at 0 hours.

If you choose not to use password protection, the default password is automatically displayed on the password screen when you start up, and you simply press Enter for access to all functions of the analyzer.

Using the Data Entry and Function Buttons

The ◄► arrow buttons select options from the menu currently being displayed on the display. When selected, the option blinks.

When the selected option includes a modifiable item, the ▲▼ arrow buttons can be used to increment or decrement that modifiable item.

The Enter button is used to accept any new entries on the VFD screen.

The Escape button is used to abort any new entries on the VFD screen that are not yet accepted by use of the Enter button.

Teledyne Analytical Instruments

8

Page 10
Image 10
Teledyne 3000TA-XL-EU quick start Default Configuration, Using the Data Entry and Function Buttons, Auto Ranging on