Quickstart Guide Trace Oxygen Analyzer
Teledyne Analytical Instruments 14
Press the Alarm button on the front panel to enter the Alarm function. Make sure that
AL–1 is blinking.
AL—1 AL—2
Choose Alarm
Using the ◄► arrow keys to move the blinking over to AL–1. Then press Enter.
AL—1 1000 ppm HI
Dft—N Fs—N Ltch—N
Five parameters can be changed on this screen:
Value of the alarm setpoint, AL–1 #### ppm (oxygen); value can be set from 0.00
to 10,000.00 ppm
Out-of-range direction, HI or LO
Defeated? Dft–Y/N (Yes/No)
Failsafe? Fs–Y/N (Yes/No)
Latching? Ltch–Y/N (Yes/No).
To change any of the alarm parameters, use the ◄► arrow keys to move the blinking
over to the specific parameter.
Use the ▲▼ arrow keys to change the value of the parameter. Holding down the key
speeds up the incrementing or decrementing.
Once the parameters for alarm 1 have been set, press Alarms again, and repeat this
procedure for alarm 2 (AL–2).
To reset a latched alarm, go to Dft– and then press either two times or two
times. (Toggle it to Y and then back to N.)
—or—
Go to Ltch– and then press either two times or two times. (Toggle it to N and
back to Y.)
Set The Range Function
To set the 3 programmable ranges:
Enter the range function mode by pressing the Range button on the front panel.
L—1 M—10
H—100 Mode—AUTO