2.1.1.1. AC configuration – Internal Pump (JP7)

Relay PCA 04523

Addendum to E-Series Operator Manuals

2.1.1.1. AC configuration – Internal Pump (JP7)

AC power configuration for internal pumps is set using Jumper set JP7 (see Figure 1-1 for the location of JP7).

 

 

Table 2-1:

AC Power Configuration for Internal Pumps (JP7)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LINE

LINE

JUMPER

 

JUMPER

 

 

FUNCTION

BETWEEN

 

 

POWER

FREQUENCY

COLOR

 

 

 

PINS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Connects pump pin 3 to 110 / 115 VAC power line

2 to 7

 

 

 

60 HZ

WHITE

 

 

 

 

 

Connects pump pin 3 to 110 / 115 VAC power line

3 to 8

 

 

 

 

 

 

 

 

 

110VAC

 

 

Connects pump pins 2 & 4 to Neutral

4 to 9

 

 

115 VAC

 

 

Connects pump pin 3 to 110 / 115 VAC power line

2 to 7

 

 

 

 

 

 

 

 

50 HZ1

BLACK

 

 

 

 

 

Connects pump pin 3 to 110 / 115 VAC power line

3 to 8

 

 

 

 

 

Connects pump pins 2 & 4 to Neutral

4 to 9

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

60 HZ

BROWN

Connects pump pins 3 and 4 together

1 to 6

 

 

220VAC

 

 

 

 

Connects pump pin 1 to 220 / 240VAC power line

3 to 8

 

 

 

 

 

 

240 VAC

50 HZ1

BLUE

Connects pump pins 3 and 4 together

1 to 6

 

 

 

 

 

 

 

 

Connects pump pin 1 to 220 / 240VAC power line

3 to 8

 

 

 

 

 

 

 

 

 

 

 

 

 

1 A jumper between pins 5 and 10 may be present on the jumper plug assembly, but is only functional on the M300E and has no function on the Models M100E, M200E or M400E.

110 VAC /115 VAC

 

220 VAC /240 VAC

1

6

 

1

6

2

7

 

2

7

3

 

8

3

8

4

9

 

4

9

5

10

 

5

10

Present on 50 Hz version of jumper set,

and functional for M300E but not Models M100E, M200E & M400E

Figure 2-3: Pump AC Power Jumpers (JP7)