2.1.1.3. AC Configuration –Heaters for Option Packages (JP6)

Relay PCA 04523

Addendum to E-Series Operator Manuals

2.1.1.3. AC Configuration –Heaters for Option Packages (JP6)

Both the IZS valve option or an O2 sensor options include AC heaters that maintain an optimum operating temperature for key components of those options. Jumper set JP6 is used to connect the heaters associated with those options to AC power. Since these heaters work with either 110/155 VAC or 220/240 VAC, there is only one jumper configuration.

Table 2-3: Power Configuration for Optional AC Heaters (JP6)

JUMPER

 

MODEL’S

 

JUMPER

 

HEATER(S)

 

BETWEEN

FUNCTION

COLOR

USED ON1

 

 

 

 

 

PINS

 

 

 

 

 

 

 

 

IZS Permeation Tube

M100E’s,

 

1 to 8

Common

 

M200E’s &

 

 

Heater

 

 

 

 

M400E

 

 

 

RED

 

2 to 7

Neutral to Load

 

 

 

 

 

 

O2 Sensor Heater

M100E’s &

 

3 to 10

Common

 

M200E’s

 

4 to 9

Neutral to Load

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

IZS

 

10 9

 

 

Permeation Tube

12 11

8

7

O2 Sensor

Heater

 

 

 

Heater

 

 

 

 

6

5

4

3

2

1

Figure 2-5: Typical Set Up of AC Heater Jumper Set (JP2)

2.2. Valve Control

Other than the location of the connectors and control chips on the PCA (see Figure 1-1 above), there have been no significant changes to the method by which the E-Series instruments’ various valves and valve options are controlled.