2.1.1.2. AC Configuration – Standard Heaters (JP2)

Addendum to E-Series Operator Manuals

Relay PCA 04523

2.1.1.2. AC Configuration – Standard Heaters (JP2)

Power configuration for the AC the standard heaters is set using Jumper set JP2 (see Figure 1-1 for the location of JP2).

Table 2-2: Power Configuration for Standard AC Heaters (JP2)

LINE VOLTAGE
JUMPER

COLOR

HEATER(S)
MODEL’S

USED ON1

JUMPER

BETWEEN

PINS

FUNCTION

 

 

Reaction Cell / Sample

M100E’s &

1 to 8

Common

 

 

Chamber Heaters

M200E’s

2 to 7

 

 

 

Neutral to Load

 

 

 

 

 

 

 

 

3 to 10

 

 

 

Mini Hi-Con

M200EH

Common

110 VAC / 115 VAC

WHITE

Converter

4 to 9

Neutral to Load

 

50Hz & 60 Hz

 

 

 

Moly Converter

M101E and

3 to 10

Common

 

 

 

 

M200E’s

4 to 9

Neutral to Load

 

 

 

 

 

 

 

 

 

 

 

5 to 12

 

 

 

By Pas Manifold

M200#M &

Common

 

 

M200EH

6 to 11

Neutral to Load

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reaction Cell / Sample

M100E’s &

1 to 7

Load

 

 

Chamber Heaters2

M200E’s

 

 

 

220 VAC / 240 VAC

 

Hi Concentration

M100EH

3 to 9

Load

BLUE

Converter

 

 

 

50Hz & 60 Hz

 

 

 

 

Moly Converter

M101E and

3 to 9

Load

 

 

 

 

 

M200E’s

 

 

 

 

 

 

 

 

 

 

 

 

 

By Pas Manifold

M200#M &

5 to 11

Load

 

 

M200EH

 

 

 

 

 

 

 

 

 

 

 

1No AC Heaters are used in the M400E series analyzers.

Reaction Cell or Sample Chamber Heaters

Mini Hi-Con or Moly Converter Heaters

200EM/EH

By Pass Manifold

Heater

1

7

1

7

 

Reaction Cell or

2

8

2

8

 

Sample Chamber

 

Heaters

3

9

3

 

9

Mini Hi-Con or

4

10

4

 

10

Moly Converter

 

Heaters

5

11

5

11

 

200EM/EH

 

 

 

 

 

By Pass Manifold

6

12

6

12

 

Heater

110 VAC /115 VAC

220 VAC / 240 VAC

 

 

Figure 2-4: Typical Set Up of AC Heater Jumper Set (JP2)