1.2. What’s Different

Addendum to E-Series Operator Manuals

Relay PCA 04523

1.2. What’s Different

The major differences between the Relay PCA 04523 and Relay PCA 03955 are:

On instruments with internal pumps, the pump is now configured for use with 110 VAC, 60 Hz vs. 220 VAC, 50 Hz operation by a set of configuration jumpers on the Relay PCA. Previously a set of inline connectors and wiring harnesses performed this function.

The configurability of the two thermocouple inputs has been expanded to allow for both grounded and ungrounded thermocouples

A retainer plate has been added to the 045230200 version to keep the solid state AC power relays securely inserted in their sockets.

Because the board has been physically changes, locations of components such as jumper blocks, connectors and status LED’s have changed.

1.3. Relay PCA Location and Layout

Generally the relay PCA is located in the right-rear quadrant of the analyzer and is mounted vertically on the back side of the same bracket as the instrument’s DC power supplies, however the exact location of the relay PCA may differ from model to model. Please check the Getting Started chapter of your M100E/M200E/M400E operator’s manual.

Thermocouple

Status LED’s

Signal Output

(D2 through D16)

(JP5)

 

Thermocouple

 

Configuration

 

Jumpers

 

(J15)

 

TC1 Input

 

(J16)

 

TC2 Input

 

(JP7)

 

Pump AC

 

Configuration

 

Jumper

 

Watchdog

DC Power Supply

Status LED (D1)

Test Points

1.3. Relay PCA Location and Layout I2C Connector

Power

Connection

for DC

Heaters

Shutter Control

Connector

(M100E Series

Only)

Figure 1-1: Relay PCA Layout (P/N04523) Valve Control

Pump Power

Output

AC Power

IN

AC Heater

Power Output

Solid State AC Power Relays

(JP6)

Drivers

Valve Option

Control

Connector

DC Power

Distribution

Connectors

(Not Present on

P/N 45230100)(JP2)

AC Configuration Jumpers

for Optional IZS Valve

Heaters & 02Sensors

Main AC Heater

Configuration Jumpers

AC Power Output for Optional IZS Valve Heaters & 02 sensors

Figure 1-1: Relay PCA Layout (P/N04523)

2

05118 Rev B3

Page 8
Image 8