Teledyne M100E INPUT TC1 J15, Output Scale Selector 7, Type J Compensation 8, Input Gain Selector

Models: M200E and M400E M100E

1 22
Download 22 pages 18.99 Kb
Page 19
Image 19
Input Gain Selector

Relay PCA 04523

Addendum to E-Series Operator Manuals

Input Gain Selector

Output Scale Selector

Type J Compensation

Type J Compensation

Termination Selector

Input Gain Selector

1 – 11

2 – 12

3 – 13

4 – 14

5 – 15

6 – 16

 

 

 

 

 

 

Output Scale Selector 7 – 17

Type J Compensation 8 – 18

Type J Compensation 9 – 19

Termination Selector 10 – 20

TC1

TC2

Figure 2-8: Thermocouple Configuration Jumper (JP5) Pin-Outs

Table 2-5: Typical Thermocouple Settings for M100E/M200E series analyzers

TC

TERMINATION

OUTPUT

JUMPER

 

JUMPER

SCALE

BETWEE

USED ON

TYPE

TYPE

COLOR

TYPE

N PINS

 

 

 

 

 

 

 

 

 

 

 

INPUT TC1 (J15)

K

GROUNDED

5mV / °C

2 – 12

M100EH with Mini HiCon Converter

BROWN

 

 

 

4 – 14

 

 

 

 

 

 

 

 

K

 

 

2 – 12

 

 

ISOLATED

5mV / °C

4 – 14

M100EH with Mini HiCon Converter

GREY

 

 

 

5 – 15

 

 

 

 

 

 

 

 

 

 

 

 

 

 

K

ISOLATED

10mV / °C

4 – 14

M100E/M200E with Moly Converter

PURPLE

 

 

 

5 – 15

 

 

 

 

 

 

 

 

J

 

 

1 – 11

 

 

ISOLATED

10mV / °C

3 – 13

M100E/M200E with Moly Converter

RED

 

 

 

5 – 15

 

 

 

 

 

 

 

 

J

GROUNDED

10mV / °C

1 – 11

M100E/M200E with Moly Converter

GREEN

 

 

 

3 – 13

 

 

05118 Rev B3

13

Page 19
Image 19
Teledyne M100E INPUT TC1 J15, Output Scale Selector 7, Type J Compensation 8, Type J Compensation 9, Input Gain Selector