2.4.1. Thermocouple Inputs and Configuration Jumper (JP5)

Addendum to E-Series Operator Manuals

Relay PCA 04523

2.4.1. Thermocouple Inputs and Configuration Jumper (JP5)

Although the 04523 relay PCA supports two thermocouple inputs, the current M100E/M200E series analyzers only utilize one and the M400E series analyzers none. By default, this single thermocouple input is plugged into the TC1 input (J15). TC2 (J16) is currently not used. See Figure 1-1 for location of J15 and J16

Table 2-4: Thermocouple Configuration Jumper (JP5) Pin-Outs

TC INPUT

JUMPER PAIR

DESCRIPTION

FUNCTION

 

 

 

 

 

 

 

 

 

 

 

Selects preamp gain factor for J or K TC

 

1 – 11

Gain Selector

- IN = J TC gain factor

 

 

 

- OUT = K TC gain factor

 

 

 

Selects preamp gain factor for J or K TC

 

 

 

 

2 – 12

Output Scale Selector

- IN = 5 mV / °C

 

 

 

- OUT = 10 mV / °C

TC1

 

 

When present, sets Cold Junction

3 – 13

Type J Compensation

Compensation for J type Thermocouple

 

 

 

When present, sets Cold Junction

 

4 – 14

Type K Compensation

 

Compensation for K type Thermocouple

 

 

 

 

 

 

Selects between Isolated and grounded

 

 

 

 

5 – 15

Termination Selector

TC

 

- IN = Isolate TC

 

 

 

 

 

 

- OUT = Grounded TC

 

 

 

 

TC2

6 – 16

Gain Selector

 

 

7 – 17

Output Scale Selector

 

 

8 – 18

Type J Compensation

 

 

9 – 19

Type K Compensation

 

 

10 – 20

Termination Selector

 

 

Same as Pins 1 – 11 above. Same as Pins 2 – 12 above. Same as Pins 3 – 13 above. Same as Pins 4 – 14 above. Same as Pins 5 – 15 above.