Addendum to | Relay PCA 04523 |
2.4.1. Thermocouple Inputs and Configuration Jumper (JP5)
Although the 04523 relay PCA supports two thermocouple inputs, the current M100E/M200E series analyzers only utilize one and the M400E series analyzers none. By default, this single thermocouple input is plugged into the TC1 input (J15). TC2 (J16) is currently not used. See Figure
Table 2-4: Thermocouple Configuration Jumper (JP5) Pin-Outs
TC INPUT | JUMPER PAIR | DESCRIPTION | FUNCTION | |
|
|
|
| |
|
|
|
| |
|
|
| Selects preamp gain factor for J or K TC | |
| 1 – 11 | Gain Selector | - IN = J TC gain factor | |
|
|
| - OUT = K TC gain factor | |
|
|
| Selects preamp gain factor for J or K TC | |
|
|
| ||
| 2 – 12 | Output Scale Selector | - IN = 5 mV / °C | |
|
|
| - OUT = 10 mV / °C | |
TC1 |
|
| When present, sets Cold Junction | |
3 – 13 | Type J Compensation | |||
Compensation for J type Thermocouple | ||||
|
|
| When present, sets Cold Junction | |
| 4 – 14 | Type K Compensation | ||
| Compensation for K type Thermocouple | |||
|
|
| ||
|
|
| Selects between Isolated and grounded | |
|
|
| ||
| 5 – 15 | Termination Selector | TC | |
| - IN = Isolate TC | |||
|
|
| ||
|
|
| - OUT = Grounded TC | |
|
|
|
|
TC2
6 – 16 | Gain Selector |
|
|
7 – 17 | Output Scale Selector |
|
|
8 – 18 | Type J Compensation |
|
|
9 – 19 | Type K Compensation |
|
|
10 – 20 | Termination Selector |
|
|
Same as Pins 1 – 11 above. Same as Pins 2 – 12 above. Same as Pins 3 – 13 above. Same as Pins 4 – 14 above. Same as Pins 5 – 15 above.
12 | 05118 Rev B3 |