11-2
11 SETTING ATMOSPHERIC CORRECTION
lHow to Set the Atmospheric Correction Value Directly
Measure the temper ature and air pres sure to find atmos pheric correction value (pp m) from the c hart or
correction formula.
[Example] Atmospheric correction value
The display returns STAR key menu.
*1) Range : Temp. -30.0 °C to +60.0 °C (0.1 °C step)
Pres. 560.0 to1066.0hPa (0.1hPa step) , 420.0 to 800.0mmHg (0.1mmHg step)
16.5 to 31.5inHg (0.1inHg step)
*2) When the atmospheric correction value, which is calculated from the input temperature and pressure
values, exceeds the range ± 999.9 ppm, the operating procedure returns to step 3 automatically. Input
values again.
Operating procedure Operation Display
1Press STAR (H) key.
Press the [F6] key twice to get the page three. [H]
[F6] twice
2Press the [F2] key.
Current setting value is displayed. [F2]
3Move the cursor (>) to ppm setting by
pressing [F5]() key. Move
cursor
4Enter atmospheric correction value and pr ess
[ENT] key. *1) Enter ppm
[ENT]
The display returns previous mode.
*1) Input range : -999.9mm ~ +999.9mm, 0.1mm step
2003-10-10 14:30:40
3
20.0 °C 5.1ppm
-2.0 mm
1012.0hPa
5.1ppm
-2.0 mm