9-10
9 CHECK AND ADJUSTMENT
7Collimate target A. Collimate
A
(Reverse)
8Press [F6](SET)key.
Repeat the procedures in step 7 and 8 so
that the count of measured times matches to
the one in FACE(1).
* 2),3),4)
The title display will be shown automatically.
[F6]
9Collimate target B (more than ±10° from the
level ) in reverse telescope setting
(FACE(2)). *5)
Collimate
B
(Reverse)
10 Press [F6](SET)key. *1) [F6]
11 Turn the telescope in normal telescope
setting(FACE(1)). Turn
telescope
12 Collimate target B. Collimate
B
(Normal)
13 Press [F6](SET) key. Repeat the procedures
in step 12 and 13 so that the count of
measured times matches to the one in
FACE(2).
Then the display returns to main menu.
[F6]
*1) It is able to get the average value from 1 to 10 measurements. To get the average, repeat the
procedures in steps 4, 5 or 9, 10. The measured times is counted in the second line of display.
*2) The compensation values of 1) Error of vertical axis (X,Y tilt sensor offset),
2) Collimation error, and 3) Error of vertical angle 0 datum will be set and memorized internally.
*3) The operating procedure steps to set compensation value of 4) Error of horizontal axis.
*4) Pressing [F1](SKIP) key enables to set next step without changing the last compensated value.
*5) Pressing [F1](SKIP) key makes end of setting without changing compensation value.
FACE 2
V:270°04'20"
SKIP SET
LEVEL±0 5/5
(B)HORIZONTAL Axis
FACE 2
V:270°04'20"
SKIP SET
LEVEL±10° /0
FACE 2
V:270°04'20"
SKIP SET
LEVEL±10° /5
FACE 1
V: 69°58'30"
SKIP SET
LEVEL±10° 5/5
Complete