VXI Technology, Inc.
REMOTE POWER-ON OPTION
If the CT-100C mainframe is to be installed in a remote location, it is possible to apply and remove power from the unit via the remote power pins located on connector J201. To utilize the remote power feature, set the front panel switch to the standby position. Remove the power cord, then remove the power supply from the chassis and place the jumper on the power interface board to the 2 – 3 position (see Figure 2-1for power supply removal instructions). The power interface board will be located inside the mainframe at the top right. Doing this will disable the front panel power switch. Pins 5 and 18 can now be used to remotely control the powering of the chassis. The circuit is diagramed below:
Jumper Location
Pin 13
Power
Supply
Pin 25
Switch
Power Interface Board
Power Interface Board
FIGURE 1-1: REMOTE POWER-ON SWITCH WIRING