2.4 Optional Terminal Wiring

Wiring Procedure

1.Turn OFF the recorder and remove the terminal cover.

2.Connect the input signal wires to the terminals.

Philips screwdriver

Crimp-on lug with insulation sleeves

3.Replace the terminal cover and fasten it with screws. The proper torque for tightening the screws is 0.6 N-m.

Alarm output

Remote control input

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FAIL/chart end output

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Relay contact input

Transistor input

Relay contact output

(Voltage-free contact)

(Open collector)

 

 

 

 

1

2

3

1

2

 

 

 

 

3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NO

C NC

4 5 C

4 5 C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Before Using the Recorder

Note

To reduce noise, use a shielded cable for the wiring of the remote control input terminals.

Connect the shield to the ground terminal of the recorder.

• Relay Contact Output Specifications

Output type:

Relay

Contact rating:

250 VAC (50/60 Hz)/3 A, 250 VDC/0.1 A (resistive load)

Dielectric strength:

1500 VAC at 50/60 Hz for one minute (between output

 

terminals and the ground terminal)

• Relay Contact Input/Transistor Input Specifications

Input signal:

Voltage-free contact: Contact closed at 200 or less and

 

contact open at 100 kor greater

 

• Open collector: 0.5 V or less (30 mADC) when turned ON,

 

leakage current of 0.25 mA or less when turned OFF

Input format:

Photocoupler isolation (shared common)

Dielectric strength:

500 VDC for one minute between input terminals and the

 

ground terminal

2-11

Page 55
Image 55
Omega RD100B Wiring Procedure, Relay Contact Output Specifications, Relay Contact Input/Transistor Input Specifications