12.6 General Specifications

Standard Performance

Item

Specifications

 

 

 

 

 

Measurement and recording accuracy

 

 

 

 

 

 

 

The following specifications apply to operation of the recorder under standard operation conditions:

 

 

Temperature:

23 ± 2°C

 

 

 

 

 

 

Humidity:

55% ± 10%RH

 

 

 

 

Power supply voltage:

90 to 132 or 180 to 264 VAC

 

 

 

 

Power supply frequency: 50/60 Hz ± 1%

 

 

 

 

Warm-up time:

At least 30 minutes.

 

 

 

 

Other ambient conditions such as vibration should not adversely affect the operation

 

 

 

 

 

 

 

 

 

Input Type

Range Type

Measurement (Digital Display)

 

Recording (Trend)

 

 

 

 

 

 

 

 

Measurement Accuracy

Highest Res.

Recording Accuracy

Resolution

 

 

 

 

DC voltage

20 mV

 

 

10 V

 

Measurement

Pen model:

 

 

60 mV

 

 

10 V

 

accuracy

Dead band 0.2%

 

 

 

 

 

 

 

± (0.3% of

of recording span

 

 

200 mV

 

 

100 V

 

 

 

 

 

 

recording span)

Dot model:

 

 

2 V

±(0.1% of rdg + 2 digits)

1 mV

 

 

 

 

 

Resolution 0.1 mm

 

 

6 V

 

 

1 mV

 

 

 

 

 

 

 

 

 

 

 

20 V

 

 

10 mV

 

 

 

 

 

50 V

±(0.1% of rdg + 2 digits)

10 mV

 

 

 

 

DC voltage

1-5 V

±(0.1% of rdg + 2 digits)

1 mV

 

 

 

 

TC

R

±(0.15% of rdg + 1)

0.1°C

 

 

 

 

 

S

except R,S: 0 to 100°C, ±3.7°C

 

 

 

 

 

(excludes

B

100 to 300°C, ±1.5°C

 

 

 

 

 

RJC

 

B: 400 to 600°C, ±2°C

 

 

 

 

 

accuracy)

 

Accuracy not warranted for

 

 

 

 

 

 

 

values less than 400°C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

K

±(0.15% of rdg + 0.7°C)

 

 

 

 

 

 

 

except ±(0.15% of rdg + 1°C)

 

 

 

 

 

 

 

for –200 to –100°C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E

±(0.15% of rdg + 0.5°C)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

J

±(0.15% of rdg + 0.5°C)

 

 

 

 

 

 

T

except ±(0.15% of rdg + 0.7°C)

 

 

 

 

 

 

 

for J: –200 to –100°C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

N

±(0.15% of rdg + 0.7°C)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

W

±(0.15% of rdg + 1°C)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

L

±(0.15% of rdg + 0.5°C)

 

 

 

 

 

 

U

 

 

 

 

 

 

except ±(0.15% of rdg + 0.7°C)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

for L: –200 to –100°C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WRe

±(0.2% of rdg + 1°C)

 

 

 

 

 

RTD

Pt100

±(0.15% of rdg + 0.3°C)

 

 

 

 

 

 

JPt100

 

 

 

 

 

 

 

ON/OFF input

Level

Threshold level (2.4 V)

 

 

 

 

Contact

accuracy ± 0.1 V

 

 

 

 

 

 

 

 

1 K: ON, 100 k: OFF

 

 

 

 

 

 

 

(Parallel capacitance

 

 

 

 

 

 

 

0.01 F or less)

 

 

 

 

 

Measuring accuracy in case of linear scaling

Accuracy during scaling (digits) = measuring accuracy (digits) multiplier + 2 digits (rounded up) where the multiplier = scaling span (digits)/measuring span (digits).

Example: Assuming that

• range type:

1-5 V

• measuring span:

1.000 to 5.000 V

• scaling span:

0.000 to 2.000 Then,

Measuring accuracy = ±(0.1% 5 V + 2 digits)

=±(0.005 V [5 digits] + 2 digits)

=±7 digits

Multiplier = {2000 digits (0.000 to 2.000)}/4000 digits (1.000 to5.000) = 0.5

Accuracy during scaling = ±(7 0.5 + 2) digits = 6 digits (rounded up)

Reference junction compensation

Internal/External selectable for each channel

12

Specifications

12-19