7.6Recording by Compensating for the Pen Offset along the Time Axis (Pen Model)

This section explains the details of setting the compensation for the pen offset along the time axis (pen model).

Basic Setting mode cannot be entered when recording is in progress or when computation is in progress on models with the computation function (/M1 option).

Procedure

Changing the Settings

1.Hold down the MENU key for 3 seconds to enter Setting mode.

2.Hold down the and keys simultaneously for 3 seconds to enter Basic Setting mode.

3.Press the key or SHIFT + key to select POC and then press the key.

4.Set each item and press the key.

Use the key or SHIFT + key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up to then are cancelled, and the display returns to a higher level menu.

5.When the Setting complete screen appears, do either of the following:

To correct the setting, press the key. If you are done, press the ESC key.

Basic=POC

 

POC=Off

 

 

 

Select On to record by compensating

 

 

 

 

 

for the pen offset along the time axis.

 

 

 

 

 

The new setting takes effect.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

POC

 

 

 

ESC/? Setting complete

Applying the Changes and Returning to Operation Mode

Press the key or SHIFT + key to select End and then press the key. Press

the key to select Store and then press the key. The changes are applied, and the screen returns to Operation mode. If you select Abort and press the key, the changes are discarded, and the screen returns to Operation mode. Press the ESC key to return to the Basic= screen.

Explanation

POC (Pen Offset Compensation)

On: Enables pen offset compensation.

Off: Disables pen offset compensation.

Note

When pen offset compensation is enabled, all pens are aligned with the reference pen (pen that is furthest to the back: pen 2 on the 2-pen model, pen 3 on the 3-pen model, and pen 4 on the 4-pen model) for recording. When recording is started, only the reference pen moves until the offset is compensated. This is not a malfunction.

When pen offset compensation is enabled, an asterisk is printed by the time on the periodic printout.