9.8 Setting the Partial Expanded Recording

This section explains the details of expanding a section of the recording range.

If you change the computing equation or the recording span, set the partial expanded recording again.

Procedure

1.Hold down the MENU key for 3 seconds to enter Setting mode.

2.Press the key or SHIFT + key to select Math and then press the key.

3.Press the key or SHIFT + key to select Aux and then press the key.

4.Press the key or SHIFT + key to select Partial and then press the key.

5.Set each item and press the key.

Use the key or SHIFT + key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up to then are cancelled, and the display returns to a higher level menu.

6.When the Setting complete screen appears, do either of the following: Press the key to set other channels.

If you are done, press the ESC key.

7.Hold down the MENU key for 3 seconds to return to Operation mode.

Set=Math

Math=Aux

Aux=Partial

CH=0A-0A

Set the channel range.

First channel Last channel

Partial=Off

Expand=50

Bound=0.00

Select On to use the partial expanded recording function.

Set the boundary position as

a percentage where the recording span is assumed to be 100%.

Set the boundary value within the recording span.

ESC/?

0A-0A Channel Setting complete

The new setting takes effect.

<Related Topics>

Enabling the partial expanded recording function: Section 7.13

9-16