12.5 Specifications of Optional Functions

Item

Specifications

 

 

 

 

Measurement and recording accuracy

 

 

 

 

 

 

 

 

 

 

 

Input Type

Measurement Accuracy

Recording Accuracy

 

 

PR40-20

0 to 450°C

Not warranted

 

 

 

 

450 to 750°C

±(0.9% of rdg + 3.2°C)

 

 

 

 

750 to 1100°C

±(0.9% of rdg + 1.3°C)

 

 

 

 

1100 to 1900°C

±(0.9% of rdg + 0.4°C)

 

 

 

PLATINEL

 

±(0.25% of rdg + 2.3°C)

 

 

 

NiNiMo

 

±(0.25% of rdg + 0.7°C)

 

 

 

W/WRe26

0 to 400°C

±15.0°C or less

 

 

 

 

400 to 2400°C

±(0.2% of rdg + 2.0°C)

 

 

 

Type N(AWG14)

 

±(0.2% of rdg + 1.3°C)

Measurement

 

 

Kp vs Au7Fe

0 to 20 K

±4.5 K

accuracy

 

 

 

20 to 300 K

±2.5 K

±(0.3% of

 

 

Pt25

 

±(0.15% of rdg + 0.6°C)

recording span)

 

 

Pt50

 

±(0.3% of rdg + 0.6°C)

 

 

 

Ni100(SAMA)

 

 

 

 

 

Ni100(DIN)

 

±(0.15% of rdg + 0.4°C)

 

 

 

Ni120

 

 

 

 

 

J263*B

0 to 40 K

±3.0 K

 

 

 

 

40 to 300 K

±1.0 K

 

 

 

Cu53

 

±(0.15% of rdg + 0.8°C)

 

 

 

Cu100

 

±(0.2% of rdg + 1.0°C)

 

 

Note) No RJC for PR40-20 (fixed to 0°C)

Remote Control 5 Points (/R1)

Item

Specifications

Number of input terminals

5

 

Input type

Isolated from the main circuitry through a photocoupler, built-in isolated power supply for the input

 

terminals, and shared common

Input type and signal level

 

 

Voltage-free contact

Contact closed at 200 or less and contact open at 100 kor greater

Open collector

ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when OFF: 0.25 A or less

Signal type

Edge, level, or trigger (250 ms or more)

Operation

Perform the following control operations by applying a given signal to the remote signal input

 

terminal.

 

Start/stop recording (edge)

 

Alarm ACK (trigger)

 

• Set the clock (adjust the clock to the nearest hour with the contact input, trigger)

Time When Signal Is Input

Procedure

00 min 00 s to 01 min 59 s

Truncates the minutes and seconds. Example: 10:00:50 10:00:00

58 min 00 s to 59 min 59 s

Rounds up the minutes and seconds. Example: 10:59:50 11:00:00

02 min 00 s to 57 min to 59 s

No operation

 

 

Manual printout (trigger)

Message 1 to 5 printout (trigger)

Switch the chart speed (level)

Start/stop computation (/M1 option) (edge)

Reset computation (/M1 option) (trigger)

12-16