Omega RD100B Setting the Input range, Selectable Range of Input Range, Span Left, and Span Right

Models: RD100B

1 234
Download 234 pages 60.28 Kb
Page 96
Image 96

5.1 Setting the Input range

Description

Selectable Range of Input Range, Span Left, and Span Right

The input range, span left, and span right can be set in the range shown below. Span left and span right cannot be set to the same value.

DC voltage (Volt)

Range Type

Selectable Span Range

 

20 mV

–20.00 to 20.00 mV

 

 

 

 

60 mV

–60.00 to 60.00 mV

 

 

 

 

200 mV

–200.0 to 200.0 mV

 

 

 

 

2 V

–2.000 to 2.000 V

 

 

 

 

6 V

–6.000 to 6.000 V

 

 

 

 

20 V

–20.00 to 20.00 V

 

 

 

 

50 V

–50.00 to 50.00 V

 

 

 

 

Thermocouple (TC)

 

 

 

 

 

Range Type

Selectable Span Range (°C)

Selectable Span Range (°F)

 

 

 

R

0.0 to 1760.0 °C

32 to 3200 °F

 

 

 

S

0.0 to 1760.0 °C

32 to 3200 °F

 

 

 

B

0.0 to 1820.0 °C

32 to 3308 °F

 

 

 

K

200.0 to 1370.0 °C

328 to 2498 °F

 

 

 

E

200.0 to 800.0 °C

328.0 to 1472.0 °F

 

 

 

J

200.0 to 1100.0 °C

328.0 to 2012.0 °F

 

 

 

T

200.0 to 400.0 °C

328.0 to 752.0 °F

 

 

 

N

0.0 to 1300.0 °C

32 to 2372 °F

 

 

 

W

0.0 to 2315.0 °C

32 to 4199 °F

 

 

 

L

200.0 to 900.0 °C

328.0 to 1652.0 °F

 

 

 

U

200.0 to 400.0 °C

328.0 to 752.0 °F

 

 

 

WRe

0.0 to 2400.0 °C

32 to 4352 °F

 

 

 

RTD

 

 

 

 

 

Range Type

Selectable Span Range (°C)

Selectable Span Range (°F)

 

 

 

PT (Pt100)

200.0 to 600.0 °C

328.0 to 1112.0 °F

 

 

 

JPT (JPt100)

200.0 to 550.0 °C

328.0 to 1022.0 °F

 

 

 

RTD

 

 

 

 

 

Range Type

Selectable Span Range (°C)

Selectable Span Range (°F)

 

 

 

PT (Pt100)

200.0 to 600.0 °C

328.0 to 1112.0 °F

 

 

 

JPT (JPt100)

200.0 to 550.0 °C

328.0 to 1022.0 °F

 

 

 

/N1 Option

Selections

Input Type

Cu1

Cu10 (GE)

Cu2

Cu10 (L&N)

Cu3

Cu10 (WEED)

Cu4

Cu10 (BAILEY)

Cu5

Cu10: a = 0.00392 at 20°C

Cu6

Cu10: a = 0.00393 at 20°C

Cu25

Cu25*: a = 0.00425 at 0°C

/N3 Option

Selections

Input Type

Note

PR

PR40-20

Thermo

PLATI

PLATINEL

couple

NiMo

NiNiMo

 

W/WRe

W/WRe26

 

N2

Type N (AWG14)

 

Kp

Kp vs Au7Fe

 

Pt4

Pt25

RTD

Pt3

Pt50

 

Ni1

Ni100 (SAMA)

 

Ni2

Ni100 (DIN)

 

Ni3

Ni120

 

 

 

 

 

J263

J263*B

 

Cu8

Cu53

 

Cu10

Cu100: a = 0.00425 at 0°C

 

5-2

Page 96
Image 96
Omega RD100B Setting the Input range, Selectable Range of Input Range, Span Left, and Span Right, N1 Option, Input Type