5.2Setting the Alarm

Explanation

Channel Range

The only case when an alarm can be set on multiple channels simultaneously is when the channels are set to the same range type (for example channel 01 and 02 are set to 2 V range). For channels on which scaling is set, the channels must be set to the same range type, same span values, and same scaling values.

Alarm Type

Symbol

Name

Note

H

High limit alarm

 

 

 

 

L

Low limit alarm

 

 

 

 

h

Difference high limit alarm

Can be specified on channels set to

 

 

delta computation.

 

 

 

l

Difference low limit alarm

Can be specified on channels set to

 

 

delta computation.

 

 

 

R

High limit on rate-of-change alarm

 

 

 

 

r

Low limit on rate-of-change alarm

 

 

 

 

T

Delay high limit alarm*

 

 

 

 

t

Delay low limit alarm*

 

 

 

 

*Selectable only when the alarm delay function is enabled. See section 7.15.

Note

On channels set to delta computation, alarms can be detected on the values illustrated in the figure below.

Alarm on a channel set to delta computation

 

 

Alarm

 

 

Alarm

 

 

H, L, R, r, T, and t

 

 

h and l

Input value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measured value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measured value on the reference channel

If high limit on rate-of-change alarm or low limit on rate-of-change alarm is specified, set the interval over which the rate-of-change is calculated.

If delay high limit alarm or delay low limit alarm is specified, set the delay time.

<Related Topics>

Setting the interval of the high limit on rate-of-change alarm or low limit on rate-of- change alarm: Section 7.1

Setting the delay time of delay high limit alarm or delay low limit alarm: Section 6.10

Enabling the alarm delay function: Section 7.15

Setting the alarm hysteresis: Section 7.1

Alarm Value

High Limit Alarm/Low Limit Alarm and Delay High Limit Alarm/Delay Low Limit Alarm

The following values can be specified.

For DC voltage, TC, and RTD input: Values within the measurable range in the

specified range (example: –2.000 to 2.000 V for the 2 V range).

For ON/OFF input (DI): 0 or 1.

For linear scaling (1-5V, scaling, and square root): A value within –5 to 105% of the scaling span except within the range of –20000 to 30000 (excluding the decimal point).

5-10

Page 104
Image 104
Omega RD100B manual Setting the Alarm, Channel Range, Alarm Type, Symbol Name