Omega RD100B manual Input Signal Wiring Dot Model, Before Using, Recorder

Models: RD100B

1 234
Download 234 pages 60.28 Kb
Page 51
Image 51

2.3 Input Signal Wiring

Dot Model

 

b

+/A

–/B

 

 

Channel 1

 

Channel 2

 

Channel 3

 

Channel 4

 

Channel 5

 

Channel 6

Screw input terminal

Channel 2

+A

+/A

–/B

–/B

 

b

b

Channel 4

 

 

Channel 6

Clamped input terminal

(/H2 option)

Channel 1

Channel 3

Channel 5

2

Before Using the

1.Turn OFF the recorder and remove the terminal cover.

2.Connect the signal wires to the terminals.

Screw input terminal

Clamped input terminal

Signal wire

Crimp-on lug with insulation sleeves

Signal wire

Flat-blade

 

screwdriver

Philips screwdriver

Note

Input signal wires of diameter less than or equal to 0.3 mm may not be secured firmly for clamped terminals (/H2). Fold over the conducting section of the wire, for example, to make sure that the wire is securely connected to the clamped terminal.

3.Replace the terminal cover and fasten it with screws. The proper torque for tightening the screws is 0.6 N-m.

Recorder

2-7

Page 51
Image 51
Omega RD100B manual Input Signal Wiring Dot Model, Before Using, Recorder