6.4Setting Recording Zones for Each Channel (Zone Recording)

This section explains the details of setting the recording zone for each measurement channel.

Procedure

1.Hold down the MENU key for 3 seconds to enter Setting mode.

2.Press the key or SHIFT + key to select Aux and then press the key.

3.Press the key or SHIFT + key to select Zone and then press the key.

4.Set each item and press the key.

Use the key or SHIFT + key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up to then are cancelled, and the display returns to a higher level menu.

5.When the Setting complete screen appears, do either of the following:

Press the key to set other channels. If you are done, press the ESC key.

6.Hold down the MENU key for 3 seconds to return to Operation mode.

Set=Aux

Aux=Zone

CH=01-01

First channel Last channel

Left=0

Right=100

Set the channel range.

Set the leftmost value of the recording zone.

Set the rightmost value of the recording zone.

ESC/?

01-01 Channel Setting complete

The new setting takes effect.

Explanation

Zone

The recording zone can be set in the following range.

Leftmost position: 0 mm to 95 mm

Rightmost position: 5 mm to 100 mm

Set a larger value for the rightmost position than the leftmost position.

Set at least 5 mm for the zone width.

Note

If the zone is set to a value less than 40 mm, scale printout of periodic printout cannot be

executed.

6-4

Page 112
Image 112
Omega RD100B manual Setting Recording Zones for Each Channel Zone Recording