9.15 Changing the Type of Reported Data Printed in Periodic Printout

Explanation

Mode

Sets the type of report data to be printed.

AVE: Prints the average value over the interval.

MIX: Prints the minimum, maximum, and average values over the interval.

SUM: Prints the sum value over the interval. If SUM is selected, set the sum scale.

MIN: Prints the minimum value over the interval.

MAX: Prints the maximum value over the interval.

INST: Prints the instantaneous value.

SUM scale

Set this item only when SUM is selected.

For the definition of the term “sum scale,” see page 7-14.

Off:

Simply sums the computed values.

/s:

Sums by converting the computed values to a value over 1 second.

/min:

Sums by converting the computed values to a value over 1 minute.

/h:

Sums by converting the computed values to a value over 1 hour.

/day:

Sums by converting the computed values to a value over 1 day.

<Related Topics>

Setting the periodic printout interval: Section 7.8

9-26