9.2 Setting the Computing Equation

The computing equation cannot be entered when the computation is in progress. If you change the computing equation or the recording span, set the alarm and the partial expanded recording again.

Procedure

1.Hold down the MENU key for 3 seconds to enter Setting mode.

2.Press the key or SHIFT + key to select Math and then press the

key.

3 Press the key with Formula shown on the screen.

4.Set each item and press the key.

Use the key or SHIFT + key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC/? key in the middle of the operation, the settings entered up to then are cancelled, and the display returns to a higher level menu.

5.When the Setting complete screen appears, do either of the following: Press the key to set other channels.

If you are done, press the ESC/? key.

6.Hold down the MENU key for 3 seconds to return to Operation mode.

Set=Math

Math=Formula

CH=0A-0A

Set the channel range.

First channel Last channel

Mode=Off

f.=01

Span_L= -200.00

Span_R= 200.00

ESC/?

0A-0A Channel

Setting complete

Select On to use the computing equation.

Set the equation.

Set the left span value.

Set the right span value.

The new setting takes effect.

Setting the Computing Equation

The CHARACTER, , and keys are used to set the computing equation.

Use the key or SHIFT + key to select the position for entering a character. Use the CHARACTER key or SHIFT + CHARACTER key to select the character type.

Use the key or SHIFT + key to select the character.

9-2

Page 159
Image 159
Omega RD100B manual Setting the Computing Equation, First channel Last channel