7.17 Initializing the Settings

This section explains the details of initializing the recorder settings to their factory default.

Be careful, because all settings except the date/time setting in the recorder will be initialized.

Basic Setting mode cannot be entered when recording is in progress or when computation is in progress on models with the computation function (/M1 option).

Procedure

1.Hold down the MENU key for 3 seconds to enter Setting mode.

2.Hold down the and keys simultaneously for 3 seconds to enter Basic Setting mode.

3.Press the key or SHIFT + key to select Initialize and then press the key.

4.Press the key to select the settings to be initialized and then press the key.

Set:

Initializes the settings in Setting mode.

Setup+Set:

Initializes the settings in Basic Setting mode and Setting mode.

5.To execute the initialization, press the key to select Yes and then press the key.

The settings are initialized, and the recorder returns to Operation mode. To cancel the initialization, press the ESC key The screen returns to the

Basic=Initialize screen.

If you press the key to select No and press then the key, nothing happens.

Basic=Initialize

7

Setup Operations

Mode=Setup+Set

Are you sure?=No

ESC/?

Explanation

Items That Are Not Initialized

Select the range to be initialized.

Select Yes and press to execute initialization and return to Operation mode.

for Changing/Adding Functions

The following items are not initialized.

Adjustment value of the pen position and the adjustment value of the printer carriage position (Basic Setting mode)

Date/Time (Setting mode)

(Basic Setting Mode)

7-27