11.4 Calibrating the Recorder

Temperature Measurement When Using an RTD

The resistance of three lead wires must be equal.

/b

/B +/A

 

 

 

Decade resistance box

Input terminals

 

 

Temperature Measurement When Using a Thermocouple

 

Copper wires

Thermocouple wires or TC extension wires

+

 

Copper

DC voltage standard

wires

+

Thermocouple

Input terminals

wires

(0°C standard temperature device)

RJC of TC input

As the measurement terminal of the recorder is generally at room temperature, the actual output of the thermocouple is different from the values given on the thermoelectromotive force table based on 0°C. The recorder performs compensation by measuring the temperature at the input terminal and adding the corresponding thermoelectromotive force to the actual output of the thermocouple. Therefore, when the measurement terminal is shorted (equivalent to the case when the detector tip is 0°C), the measured value indicates the temperature of the input terminal.

When calibrating the recorder, this compensation voltage (thermoelectromotive force of

0°C reference corresponding to the input terminal temperature) must be subtracted from the output of the DC voltage standard before application. As shown in the figure, by using the 0°C standard temperature device to compensate the reference junction at 0 °C, you can input the thermoelectromotive force of 0 °C reference from the DC voltage standard and perform the calibration.

11

Maintenance

11-5

Page 197
Image 197
Omega RD100B manual Temperature Measurement When Using a Thermocouple, RJC of TC input, Maintenance 11-5